Temperature Compensation for Concentration of NO2 Based on DOAS
Temperature has an effect on the accuracy of measurement results of NO2 by using differential optical absorption spectroscopy(DOAS)technology to monitor the concentration of polluted gases in air in real time.In this paper,a new method of temperature compensation is proposed,in which the integral value of differential absorption cross section is linearly fitted with the temperature.Fitting functions in selected five wave-bands are used to respectively compensate,and the results are compared.The experimental results show that the relative error after compensation is less than 1.67%,which can significantly improve the accuracy of NO2 concentration measurement.
Gas detectors Temperature compensation Integration Absorption spectroscopy Differential optical absorption spectroscopy
Jingfang Zhao Yongcai Guo Fenglian Fan Yanping Wang Fang Chen
Key Laboratory of Optoelectronic Technology and Systems of Ministry of Education,School of Optoelectronic Engineering,Chongqing University,Chongqing 400044,China
国际会议
大连
英文
309-314
2019-03-29(万方平台首次上网日期,不代表论文的发表时间)