DETERMINATION OF AIR-KERMA STANDARD IN MAMMOGRAPHY X-RAY
Mammography x-rays standard is basis of improving the quality of mammary machine and reduce the dose of women and staff.The molybdenum target x-rays machine is used as the radiation source,and free-air cylindrical chamber is designed for detecting photon.Refer to international standards file four standard radiation qualities are established.The Correction factors which include air attenuation,scattered radiation,fluorescence,wall transmission,electron loss and so on are studied in the four radiation quality.The uncertainty in absolute measurement of air-kerma include air volume measurement of ionization chambers,current measurement,measurement of temperature and pressure,measurement of correction factors,physical constants and orientation of the reference distance.The relative combined standard uncertainty is 0.24%.
Metrology Free-air Chamber Air-kerma Correction factor Degree of uncertainty
S.M.Guo Y.Liu J.J.Wu J.Zhang
National Institute of Metrology,Beijing,China