Comparison of Regression Analysis and Transfer Function in Estimating the Parameters of Central Pulse Waves from Brachial Pulse Wave
This study analyzed ascending branch slope (A_slope), dicrotic notch height (Hn), diastolic area (Ad) and systolic area (As) diastolic blood pressure (DBP), systolic blood pressure (SBP), pulse pressure (PP), subendocardial viability ratio (SEVR), waveform parameter (k), stroke volume (SV), cardiac output (CO), and peripheral resistance (RS) of central pulse wave invasively and non-invasively measured. Invasively measured parameters were compared with parameters measured from brachial pulse waves by regression model and transfer function model. Accuracy of parameters estimated by regression and transfer function model, was compared too. Findings showed that k value, central pulse wave and brachial pulse wave parameters invasively measured, correlated positively. Regression model parameters including A_slope, DBP, SEVR, and transfer function model parameters had good consistency with parameters invasively measured. They had same effect of consistency. SBP, PP, SV, and CO could be calculated through the regression model, but their accuracies were worse than that of transfer function model.
Pulse wave analysis Regression analysis Data accuracy
Chai Rui Xu Li-sheng Yao Yang Hao Li-ling Qi Lin
Sino-Dutch Biomedical and Information Engineering School,Northeastern University,Shenyang,China
国际会议
第十六届世界医药健康信息学大会((MEDINFO2017)、第二届世界医药健康信息学华语论坛(WCHIS 2017)、第15届全国医药信息学大会(CMIA 2017)
苏州
英文
573-577
2017-08-21(万方平台首次上网日期,不代表论文的发表时间)