Highly sensitive detection of hexavalent chromium utilizing a sol-gel/carbon nanotube modified electrode
A pyridine-functionalized thin film has been fabricated to selectively preconcentrate Cr(Ⅵ) anions for electrochemical detection in the 5-300 μg L-1 range.Glassy carbon electrodes were modified through physical deposition of single-walled carbon nanotubes (SWNTs) on the electrode surface,followed by electrochemical deposition of a sol-gel containing a 2-pyridine functional group.The use of SWNTs has increased sensitivity for Cr(Ⅵ) detection in aqueous solutions,providing a detection limit of 0.8 μg L-1.
Chromium detection Single-walled carbon nanotubes Glassy carbon electrode Sol-gel Thin film Iridescence
Samuel M.Rosolina Stefanie A.Bragg Ruizhuo Ouyang James Q.Chambers Zi-Ling Xue
Department of Chemistry,University of Tennessee,Knoxville,TN 37996,USA