Comparison between Data from two Different Calorimeters

  In order to analyze how to use data gotten from high thermal inertial test in vent sizing, accelerating rate calorimeter and vent sizing package 2 were employed to do test on 40% hydrogen peroxide solution.Then comparison was done between this two adiabatic data.Test temperature range was set between 40-300℃, temperature step was set at 5 ℃, test temperature sensitivity was 0.02 ℃.min-1.Hydrogen peroxide solution used in ARC was 5g, hydrogen peroxide solution used in VSP2 was 70g.Thermal inertial of ARC test was 1.34 and thermal inertial of VSP2 test was 1.09.

ARC VSP2 adiabatic correction pressure correction venting size

YAO Yadong CHENG Yongxi HE Fang

Beijing Institute of Aerospace Testing Technology, Beijing 100074, China;China Aerospace Liquid Propellant Research Center, Beijing 100074, China