会议专题

R Wave Extraction Based on the Maximum First Derivative plus the Maximum Value of the Double Search

  R-wave detection is the main approach for heart rate variability analysis and clinical application based on R-R interval.The maximum first derivative plus the maximum value of the double search algorithm is applied on electrocardiogram (ECG) of MIH-BIT Arrhythmia Database to extract R wave.Through the study of algorithms characteristics and R-wave detection method,data segmentation method is modified to improve the detection accuracy.After segmentation modification,average accuracy rate of 6 sets of short ECG data increase from 82.51% to 93.70%,and the average accuracy rate of 11 groups long-range data is 96.61%.Test results prove that the algorithm and segmentation method can accurately locate R wave and have good effectiveness and versatility,but may exist some undetected problems due to algorithm implementation.

heart rate variability R-wave detection first derivative

Wen-po Yao Wen-li Yao Min Wu Tie-bing Liu

Nanjing General Hospital of Nanjing Military Command, Nanjing 210002, Jiangsu Province, China China University of Mining and Technology, Xuzhou 221116, Jiangsu Province, China

国际会议

2015 The First International Clinical Engineering and Health Technology Management Congress (ICEHTMC)(2015首届国际临床工程与医疗技术管理大会)

杭州

英文

1-5

2015-10-21(万方平台首次上网日期,不代表论文的发表时间)