会议专题

Immunosensor Based on Screen-printed Electrodes for the Determination of Myeloperoxidase

  The production and assembling of disposable electrochemical myeloperoxidase (MPO) immunosensor,which can combine the high selectivity of immunoanalysis with the ease of screen-printed electrodes,has been carried out.In the immunosensor construction,a screen-printed electrodes (SPE) was first coated by a film of multi-wall carbon nanotubes and N,N-dimethylformamide,then the gold nanoparticles (GNPs)-Chitosan homogeneous suspension has been obtained by electrochemical deposition.And then the antibody of MPO (anti-MPO) was absorbed on the surface of GNPs-Chitosan monolayer.Finally,BSA blocked the non-specific adsorption site.The modification process was characterized by transmission electron microscope and cyclic voltammetry.The factors possibly influenced the performance of the proposed immunosensor were studied in detail.Under optimal conditions,The SPE exhibited good electrochemical behavior to the concentration of MPO from 0 to 20 U/L(r=0.98501) and gave a detection limit of 0.3 U/L.Data concerning the application of modified screen-printed electrodes to the analysis of MPO in serum samples are presented and discussed.

myeloperoxidase screen-printed electrode multi-wall carbon nanotube gold nanoparticles

Xiao Ma Cheng Yu Ling-song Lu Li-ming Mo Ke-yi Wang

Department of Medical Engineering, Hangzhou First Peoples Hospital, Hangzho1 310006, Zhejiang Provi Centralab, Hangzhou First Peoples Hospital, Hangzho1 310006, Zhejiang Province, China

国际会议

2015 The First International Clinical Engineering and Health Technology Management Congress (ICEHTMC)(2015首届国际临床工程与医疗技术管理大会)

杭州

英文

10-14

2015-10-21(万方平台首次上网日期,不代表论文的发表时间)