A novel RF resonator using microstrip transmission line for human body MRI at 3T
A square-slots loaded (SSL) radio frequency (RF)resonator using microstrip transmission line (MTL) is designed for a human body 3T MRI.The SSL RF resonator shows greater penetrated RF magnetic fields near the center of the phantom than traditional RF resonators using MTL.A multichannel coil using SSL RF resonators was also simulated and provides good parallel excitation performance.In addition,RF shimming for homogenization can be effectively controlled by adjusting the inputs of eight resonators.Numerical results were obtained using a spherical phantom and a realistic human body model at 3T to calculate B+1 fields.
Hyeok-Woo Son Young-Ki Cho Hyungsuk Yoo
School of Electronics Engineering, Kyungpook National University80 Daehakro, Buk-gu, Daegu, 702-701, Department of Biomedical Engineering, University of Ulsan Department of Biomedical Engineering, Univ
国际会议
2013 International Symposium on Antennas and Propagation(2013天线与传播国际会议)
南京
英文
1141-1144
2013-10-23(万方平台首次上网日期,不代表论文的发表时间)