Applicator for Superficial Microwave Hyperthermia Filled with Anisotropic Dielectric Media:Numerical Investigation of Design Degrees of Freedom
In this paper properties of an rectangular-waveguide applicator(test applicator)for superficial microwave hyperthermia filled with stratified isotropic and/or anisotropic dielectric media are discussed.Man-made biaxial anisotropic material is considered here.Electric field distributions in a cross-section of the test applicator filled with anisotropic material are compared with the test applicator filled completely with water and with the one partially filled with water and partially with high permittivity materials.Furthermore,for some applicators power loss densities in the virtual phantom of muscle tissue are numerically estimated.
J.Vrba, Jr. D.Vrba M.Wiewegh J.Vrba
Faculty of Biomedical Engineering,Czech Technical University in Prague,Czech Republic Faculty of Electrical Engineering,Czech Technical University in Prague,Czech Republic
国际会议
Progress in Electromagnetics Research Symposium 2014(2014年电磁学研究新进展学术研讨会)
广州
英文
2741-2745
2014-08-01(万方平台首次上网日期,不代表论文的发表时间)