Assessment of Decoupling between MRI Array Elements at 300MHz

  We numerically assessed the decoupling between magnetic resonance imaging(MRI)array elements at 300 MHz.The array was loaded by a typical human head model.A fine-tuning strategy,which mimics commonly used optimization during fabrication,was applied to arrays with inductive decoupling networks between adjacent elements.Element matching(Sxx)could also be optimized for an array lacking such decoupling networks.However,a single-resonance Sxx spectrum does not provide reliable evidence of low level array element coupling.It is not easy to define the lowest limit for element coupling that does not affect array performance.For a given array configuration and a given set of excitation modes,the power reflected by the entire array provides an excellent figure-of-merit for net array decoupling and circuit-level transmit performance.