Barker Coded Excitation Using LFM Carrier for Improving Axial Resolution in Ultrasound Imaging
Improving the axial resolution of ultrasound imaging system can have broad clinical impact.In this paper,a novel Barker coded excitation uses LFM carrier (LFM-Barker coded excitation) instead of sinusoid carrier (Sinusoid-Barker coded excitation).Pulse compression scheme of LFM-Barker coded excitation is developed that consists of LFM mismatched filter,Barker matched filter and sidelobe suppression filter.In simulation,the axial resolution is almost doubled with a slight decrease of SNR gain of less than 1 dB using LFM-Barker coded excitation compared to Sinusoid-Barker coded excitation.The –6 dB mainlobe width is 0.6 μs and 1.5 μs,respectively.Maximum range sidelobe level is observed at –35 dB.Additionally LFMBarker coded excitation is robust for frequency dependant attenuation of tissue.Simulation of B-mode images is carried out to demonstrate the advantage of LFM-Barker coded excitation system of good axial resolution.
Ultrasound imaging Barker coded excitation LFM Pulse compression Axial resolution