Investigation about Dilution Ratio Effect on X-ray Intensity

  In this paper,influence coefficient of flux (αi,f) on X-ray intensity was proposed,by which the X-ray intensity of fusion beads with different dilution ratio can be corrected to the standard value.Hence,the unknown sample of random dilution ratio can be determined by the standard working curve after corrected with αi,f.It was found that αi,f obtained by the proposed experimental method can be applied in correction of fluorescence intensity in the routine analytical work with satisfactory results.There is no necessary of exact weighing for the sample and the flux with the correction ofαi,f.Furthermore,the procedure of routine analytical work can be simplified greatly without getting the L.O.I ahead of weighing,by which the working efficiency can be improved as well.

influence efficient of flux cement X-ray fluorescence analysis