会议专题

A Comparison of Cylindricity and Parallelism Measurement Procedures between HIT UOI and PTB

  PTB and the HIT Institute UOI (Center of Ultra-precision Optoelectronic Instrument) have developed special error separation procedures for improving the measurement uncertainty of cylinder calibrations.Three glass cylinders.which were custom-made by HIT,were selected to compare the performance and characteristics of the measurement procedures.Both procedures effectively eliminate guidance errors of the measurement machines and lead to sufficiently low measurement uncertainties.

precision measurement comparison error separation cylindricity

O.Jusko J. Huang W.Gu H.Reimann

Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig, Germany Harbin Institute of Technology, No.92 West Dazhi Street, Nangang District, Harbin 150001

国际会议

2012第八届精密工程测量和仪器仪表国际研讨会(ISPEMI2012)

成都

英文

1-10

2012-08-08(万方平台首次上网日期,不代表论文的发表时间)