会议专题

Testing of the scale linearity of the corrected homodyne interferometer by the femtosecond comb and optical resonator

  The work presents a method where the change of the length of the measuring arm of the Michelson interferometer is monitored by the cavity length of the optical resonator.The resonator works as distance-to-optical frequency converter with ultimate linearity.It has locked a tunable laser to certain cavity mode.The optical frequency of the laser is heterodyned with immediate “tooth of the femtosecond comb spectrum.The results of measured scale linearity of the homodyne interferometer for corrected and non-corrected scale are presented for measuring range of the homodyne interferometer in order of one micrometer and more.

laser interferometer scale linearity femtosecond comb optical resonator

O. (C)íp M. (C)í(z)ek Z. Buchta B. Mikel J. Lazar R. (S)míd

Institute of Scientific Instruments, Academy of Sciences of the Czech Republic,Královopolská 147, Brno, 612 64, Czech Republic

国际会议

2012第八届精密工程测量和仪器仪表国际研讨会(ISPEMI2012)

成都

英文

1-6

2012-08-08(万方平台首次上网日期,不代表论文的发表时间)