Iodide Selective Electrode Based on Cobalt(Ⅱ) Complex with Dual-core Schif Base as Neutral Carrier
A new highly selective poly vinyl chloride(PVC) membrane iodide electrode which has neutral carrier of cobalt(Ⅱ) complex with symmetrical binuclear schiff base has been studied.Based on the fact we knew that ethylenediamine could react to acetylacetone and salicylaldehyde.The electrode exhibited near-Nernst response to iodide with a slope of —54.4 mV/dec over a wide concentration range (1.58×10-7~1.0×10-1mol/L) with a detection limit of 6.3×10-8 mol/L in phosphorate bufer solution with pH=2.4 at 26℃.After applied to the determination of iodide in pharmaceuticals and salts,the electrode could get satisfied results
cobalt(Ⅱ) complex with symmetrical binuclear schif base neutral carrier iodide selective electrode
Gang Xu Wenli Dong Lingyan Ren
Chongqing University of Science & Technology, Chongqing, 401331, China
国际会议
太原
英文
221-224
2012-06-23(万方平台首次上网日期,不代表论文的发表时间)