A Hybrid-structured Microfluidic Chip Developed for ATP Bioluminescence Detection
A hybrid-structured microfluidic chip integrating optical fiber sensing technology and bioluminescence chemical reaction mechanism was developed for rapid and sensitive detection of adenosine 5-triphosphate (ATP).This microfluidic chip was fabricated with five distinct layers for the successive steps of sample-enzyme reagent mixing,bioluminescence chemical reaction and scatting light collection.The experimental results demonstrated that the provided hybrid-structure improved the sensitivity,response time,reagent consumption and other properties of chip.This improvement enables the microfluidic chip to attain a sensitive ATP detection in the range from 10-9 to 10-4 M with excellent linearity (R2 =0.9967),low detection error (CV < 15%),and tiny reagent consumption (< 20 μL).These achievements thus indicated that the chip developed in this study possess the advantages such as high sensitivity,quick-response and small reagent consumption.
Microfluidic chip Bioluminescence detection ATP Fiber optics Micro mixer Biosensor
Wu Ming Zhang Zhen Yu Li Tao Chen Jian Jiang Yao Yang Lv Min Li Guang Li
State Key Laboratory of Industrial Control Technology, Zhenjiang University, Hangzhou 310027,P.R.China
国际会议
广州
英文
563-569
2012-11-16(万方平台首次上网日期,不代表论文的发表时间)