Practical Advances on the Hot Wire Technique
In this work,a numerical simulation model is proposed with the purpose of determining the minimum and maximum measuring time for the hot wire parallel technique.An alternative experimental arrangement is also proposed for materials with thermal conductivity higher than 15W/mk.In this ease,the experimental thermal transient is detected very close to surface of the hot wire,and it will be called hot wire surface technique.When using this arrangement,the time interval considered for the calculations may be several times bigger than that one corresponding to the parallel technique.Experimental results obtained with the surface technique are also checked against those ones obtained with the parallel technique.
Hot wire technique Numerical simulation Minimum and maximum measuring times
Wilson Nunes dos Santos
Federal University of S(a)o Carlos-Department of Materials Engineering-Via Washington Luiz,km 235-CP 676-Sao Carlos-SP-BRAZIL-CEP 13565-905
国际会议
The Fifth China International Conference on High-Performance Ceramics (第五届先进陶瓷国际研讨会(CICC-5))
长沙
英文
1702-1705
2007-05-10(万方平台首次上网日期,不代表论文的发表时间)