A magnetic PDMS micropump for implantable drug delivery application
This paper presents an integrated magnetic PDMS-membrane micropump for implantable drug delivery application.The micropump mainly consists of two functional PDMS layers: one integrating microchannels,a pump chamber and two planar check valves into a single layer,allowing for simple design and easy system integration,and the other covering the chamber and holding a thin electroplated permalloy piece on top for magnetic actuation.The micropump can be driven by the interaction between the ferromagnetic permalloy piece and an external electromagnet,providing a remote and wireless operation method.Test results demonstrate that this micropump is able to produce a maximal flow rate of 0.15 μL/min at the driving frequency of 2 Hz,demonstrating its ability to deliver fluid samples with a volume resolution of approximately 1 nL,promising its application in various implantable biomedical systems.
micropump magnetic PDMS drug delivery
Junhui Ni Baishao Zhan Wenwu Zhong
College of Mechanical Engineering, Taizhou University, Zhejiang, China College of Physics & Electronic Engineering, Taizhou University, Zhejiang, China
国际会议
中国微米纳米技术学会第14届学术年会、第3届国际年会暨第6届微米纳米技术“创新与产业化国际研讨与展览会(CSMNT2012 & ICMAN2012)
杭州
英文
1-6
2012-11-04(万方平台首次上网日期,不代表论文的发表时间)