Electromagnetic Compatibility:RFID and Medical Equipment in Hospitals
The objective of this study is to evaluate if some life-supporting medical electrical equipment can be affected by electromagnetic waves emitted by sources for the remote identification (RFID).In particular,we assessed the occurrence of electromagnetic interference with an in-situ testing method specifically developed,based on standard ANSI C63.18,within an intensive care unit.Some serious faults were identified,both related to the RFID sources and to power sources.
Electromagnetic Compatibility Radio Frequency Identification Essential Performance Intensive Care ME Equipment
E. Iadanza M. Chini F. Marini
Dipartimento di Elettronica e Telecomunicazioni, Università degli Studi di Firenze, Firenze, Italy Ospedale Santa Maria nuova di Firenze, Firenze, Italy
国际会议
World Congress on Medical Physics and Biomedical Engineering (2012年医学物理及生物医学工程国际会议(IFMBE))
北京
英文
732-735
2012-05-26(万方平台首次上网日期,不代表论文的发表时间)