会议专题

Wavelets in the Computed Evaluation of Mammographic Accreditation Phantom Images

  This paper presents a novel approach to the computed assessment of a mammographic phantom device.The approach shown here is fully automated and is based on the automatic selection of the region of interest,in the use of the discrete wavelet transform (DWT) and morphological operators to assess the quality of the American College of Radiology (ACR) mammographic phantom images.The algorithms developed here have succesfully scored 30 images obtained with different combinations of voltage applied to the tube and exposure and could notice the differences in the radiographs due to the different level of exposure to radiation.

mammography quality control accreditation phantom computed radiography computed image evaluation

M. Alvarez D.R. Pina S.B.Duarte R.T.F. Souza A.F. Velo A.P. Trindade J.R.A. Miranda

Univ. Estadual Paulista Júlio de Mesquita Filho-UNESP/Departamento de Física e Biofísica, Laboratóri Centro Brasileiro de Pesquisas Física - CBPF, Rio de Janeiro, Brazil

国际会议

World Congress on Medical Physics and Biomedical Engineering (2012年医学物理及生物医学工程国际会议(IFMBE))

北京

英文

811-814

2012-05-26(万方平台首次上网日期,不代表论文的发表时间)