会议专题

A Novel Method Using Camera Calibration Technique for Complete Geometric Calibration of Cone-beam CT

  Cone-beam computed tomography (CBCT) is deployed in large numbers for small animal imaging.For accurate reconstruction,geometric parameters of CBCT must be known.In this paper,a novel method using camera calibration technique for CBCT geometric calibration is proposed.Geometric relationships of X-ray source,detector and rotation axis are obtained to calculate CBCT geometric parameters.Simulation results show that accuracy of this method is improved compared with previous methods.

cone-beam computed tomography geometric calibration camera calibration technique

Mengjiao Wang Hui Ding Guangzhi Wang

Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China

国际会议

World Congress on Medical Physics and Biomedical Engineering (2012年医学物理及生物医学工程国际会议(IFMBE))

北京

英文

842-845

2012-05-26(万方平台首次上网日期,不代表论文的发表时间)