How Accurate Scatter Can Be Estimated in Nuclear Medicine Images?:A Monte Carlo Study

  The quality of single-photon emission computed tomography (SPECT) imaging is hampered by scatter,attenuation and collimator blurring,and Especially,efficient scatter correction has proved to be difficult to achieve.Objective: The objective of this article was to assess the accuracy of three wellknown methods of scatter correction in comparison to montecarlo studies.methods: Different homogenous phantoms with non-uniform activity distributions and attenuation maps were simulated by SimSET then DEW,TEW and Klein-Nishina Scatter correctoin methods applied on different projections of SPECT for scatter estimation.the result of scatter estimation by these three methods compare to scatter projections simulated.Results: Klein-Nishina scatter estimation & TEW has the least differences to simulated scatter.

scatter correction nuclear medicine images DEW TEW Klein-Nishina