会议专题

Research on the system of algae concentration fluorescence measurement

The excessive growth of algae is the main reason for water blooms , how to detect the algae concentration is an important task right now. Based on the characteristics of algae chlorophyll-a fluorescence spectra, it designs a detection device of algae concentration, analyzes the basic theory of detection method, explains the structure of this detection and test process. By the experiments and data analysis, the results show that this detection has the feature of convenient and useful. The precision is 0.lμg/L. Algae concentration and fluorescence ratio come into linear relationship obviously. The linear correlation coefficient is 0.9987. By using division circuit, the relationship between excitation and fluorescence ratio gets better.

Words-Water blooms algae concentrations fluorescence measurement logarithm division circuit

Kaiqiang Wang Wenjia Yin Yu Li Yuanze Xu

Polytechnic Engineering, Qiongzhou University, Sanya, China Automation Engineering, Northeast Dianli University, Jilin, China

国际会议

2012 Fifth International Symposium on Computational Intelligence and Design 第五届计算智能与设计国际会议 ISCID 2012

杭州

英文

523-526

2012-10-28(万方平台首次上网日期,不代表论文的发表时间)