会议专题

Design of Programmable Near Infrared Cold Light Source

In the near infrared spectroscopy, detector of Charge-coupled Device requires very high stability for light source. We design the cold light source in the measurement of near infrared spectroscopy. Cold light source consists of high-power op amp OPA549 high current constant current source and proportion integral differential (PID) control algorithm. As is shown in the experimental results, the method is better, the overshoot is very small, the stability is excellent in the process of soft-start. While light intensity was able to meet the requirements of nearinfrared instruments to measure.

Near infrared Halogen Constant current source Soft-start Cold light source

Rui Wang Qingbo Fu Hui Suo Chun Zhao Biao Wang

State Key Laboratory on Integrated Optoelectronics College of Electronic Science and EngineeringJili Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences,Changchun 130

国际会议

2011 International Conference on Electronics and Optoelectronics(2011电子学与光电子学国际会议 ICEOE 2011)

大连

英文

140-143

2011-07-29(万方平台首次上网日期,不代表论文的发表时间)