Difference-Frequency-Based Laser Spectrometer for Trace CH4 Detection
A laser spectrometer based on difference-frequency generation in periodically poled LiNbO3 (PPLN) has been developed. The mid-infrared laser source was pumped by complementally two commercial narrow linewidth diode laser, to produce about lμm of output power in the tuning region from 3.2um to 3.7μm. The spectrometer was used for trace CH4 measurement, a CH4 absorption line at 3028.751 cm-1 of V3 band was chosen as this line is relatively free of interference by other gas species. The selected absorption line was verified using a single-pass cell detection. The detection sensitivity is estimated to be 0.8ppm for 10cm optical path length at atmospheric pressure, and is limited mainly by the etalon noise of optical surfaces.
Difference-frequency generation (DFG) Methane(CH4) Mid-infrared laser spectroscopy spectrometer)