Cardiovascular Multi-parameter Monitoring System during Surgery

Cardiovascular Monitoring11, which is important evidence used to analyze the therapeutic efficacy, plays a significant role in the operation.A multiparameter monitoring system is introduced in this paper.The system not only detects initial physiologic signals like photoplethysmographic pulse signal, blood pressure and electrocardiogram signal, but also extracts cardiovascular parameters, including the amplitude of photoplethysmogram, area ratio, pulse beat interval, pulse decay time constant, etc.The operating principle of the system, hardware composition, a flow chart of software module, direction of data flow and algorithm for extracting parameters are introduced.Finally, a validation clinical experiment was undertaken, and results confirmed that the system realized real-time monitoring of cardiovascular parameters, which reflected the variation of cardiovascular system during surgery and could assist doctors with drug administration.
multi-parameter monitoring system surgery therapeutic efficacy drug administration
Jing Liu Shuming Ye Hang Chen Xuan Wang Xiuquan Fu
Department of Biomedical Engineering and Instrument Science,Hangzhou,China
国际会议
厦门
英文
646-650
2011-07-08(万方平台首次上网日期,不代表论文的发表时间)