会议专题

Suture Density Study of the Bioprosthetic Heart Valve

In order to investigate the effect of the suture density on the mechanical properties of bioprosthetic heart valve, we establish the spherical leaflets models via computer aided design. Finite element analysis is used simulate the mechanical performance of bioprosthetic heart valve when leaflet is closed. The analysis results show that suture density has a significant effect to the stress state of valve, which may lead not only to different stress peak values, but also to different stress distributions. High or low density suture isnt appropriate for long-term durability of the bioprosthetic heart valve. Suture density tests are needed to perform on the valves to find satisfactory suture density for better mechanical properties of the bioprosthetic heart valve.

Bioprosthetic heart valve suture density Finite element analysis computer aided design Valve performance

Yihao Zheng Xin Ye

School of Mechanical Engineering, Shandong University, Jinan 250061, P. R. China

国际会议

2011 International Conference on Mechatronics and Applied Mechanics(2011年机电一体化与应用力学国际会议 ICMAM 2011)

香港

英文

935-939

2011-12-27(万方平台首次上网日期,不代表论文的发表时间)