会议专题

The Efficient Realization of B Mode Ultrasonic Digital Scan Conversion Based On Radix-4 CORDIC Algorithm

Coordinate transform is the core technology in Digital Scan Conversion (DSC) of B-mode ultrasonic, and its speed mostly determines the frame rate of Bmode imaging. In the traditional coordinate transform process, ultrasonic echo data is converted from polar coordinate to rectangular directly; the process is time-consuming and the data is unsuitable for posttreatment In this paper, we employed the CORDIC Algorithm in coordinate transform. We compared the conventional radix-2 CORDIC algorithm with the improved radix-4 CORDIC algorithm and found the later saved half iterations at the same frequency. Whats more, radix-4 CORDIC algorithm had smaller hardware consumption when using the pipelined structure. These results revealed that radix-4 CORDIC is an excellent method for high-speed DSC of B-mode ultrasonic.

coordinate transform DSC radix-4 CORDIC algorithm FPGA architecture

Peng Zeng Fanshui Lu Xingqun Zhao

Department of Biomedical Engineering, Southeast University, Nanjing Jiangsu, 210096, China

国际会议

2011 4th International Congress on Image and Signal Processing(第四届图像与信号处理国际学术会议 CISP 2011)

上海

英文

2714-2717

2011-10-15(万方平台首次上网日期,不代表论文的发表时间)