A Novel Method for MTF Measurement of Digital Radiographic Systems
The exact determination of the modulation transfer function (MTF) of X-ray digital imaging system is a persistent problem. We present a practical method for measuring the MTF curve of X-ray linear flat panel detector. The method is based on the noise transfer theory of cascaded linear system. The scale factor between noise power spectrum (NPS) and MTF should be obtained firstly. Accurate NPS estimation of system noise (quantum noise and additive noise) by background trends removal and MTF values measurement by standard bar pattern method were required in this method.
Modulation transfer function noise power spectrum cascased linear system standard bar pattern
Zhongxing Zhou Feng Gao Huijuan Zhao Lixin Zhang GuoHui Li
School of Precision Instrument and Optoelectronics Engineering,Tianjin University,Tianjin, China TEDA Orking Hi-Tech Co. LTD The Tianjin University Science and Technology Park Tianjin, China
国际会议
上海
英文
62-65
2011-10-15(万方平台首次上网日期,不代表论文的发表时间)