会议专题

Study of Time Of Flight capability with Lutetium Fine Silicate and Multi-Pixel Photon Counter

We have investigated the timing resolution and spatial resolution of a pair detector which consists of a MPPC and an LFS scintillator for the next generation positron emission tomography (PET) system. A combination of these new devices allows us to improve the timing resolution for Time Of Flight (TOF) - PET. The timing resolution was measured to be 96 ps in FWHM extracted from the photon detection time difference of two detectors. This indicates significant progress of the development of TOF capability for the PET system.

TOF-PET timing resolution LFSJLYSO MPPC

Makoto Yamazaki Tohru Takeshita Yoji Hasegawa

Interdisciplinary Graduate School of Science and Engineering, Shinshu University,Matsumoto, Japan Faculty of Science, Department of Physics, Shinshu University,Matsumoto, Japan

国际会议

2011 4th International Conference on Biomedical Engineering and Informatics(第四届生物医学工程与信息学国际会议 BMEI 2011)

上海

英文

508-510