会议专题

Research on Temperature Measuring Positions Selection and Verification for Polymerase Chain Reaction Instruments

Specific fragments of DNA in vitro can be rapidly amplified after several thermal cycles of the test samples by taking Polymerase Chain Reaction technology. The good performance of temperature control in thermal cycles is critical to successful PCR reaction. However, in order to achieve good performance of temperature control and guarantee the temperature qualities of PCR instruments, it is crucial to precisely measure the temperature field of sample block by selecting appropriate temperature measuring positions on the sample block. Simulation in finite element analysis (FEA) is carried out by using software ABAQUS 6.9 in this paper in order to select appropriate temperature measuring positions on the sample block, meanwhile, the experiment concerned is conducted . The experimental results indicate that the temperature field distribution of sample block is able to be fully represented and illustrated through 13 selected temperature measuring positions.

PCR temperature measuring position FEA experiment

Kewei Hu Zhangwei Chen Jing Huang

State Key Laboratory of Fluid Power and Control Zhejiang University Hangzhou, 310027, P.R. China

国际会议

2011 4th International Conference on Biomedical Engineering and Informatics(第四届生物医学工程与信息学国际会议 BMEI 2011)

上海

英文

1171-1175

2011-10-15(万方平台首次上网日期,不代表论文的发表时间)