Thickness Dependence on Sensitivity of Guided- Mode-Resonance Biosensors
In this letter, the waveguide thickness dependence on the sensitivity of the guided-mode resonance (GMR) biosensor has been explored by simulation. It is clarified the relationship between the resonant mode number and the the waveguide thickness of GMR biosensors. By increasing the waveguide thickness can not only introduce more higher-order guided modes but also reduce both the corresponding resonant bandwidth and the resonant peak wavelength shift. Its sensitivity can improved by carefully design the GMR structure and selected the resonant mode characteristics. Carefully design the GMR structure and selected the resonant mode, the sensitive of the two-layer GMR bio-sensor will be well improved.
guided-mode resonance GMR biosensor RCWA
Jiann-Hwa Lue Ting-Jou Ding Tsung-Hsun Yang
Department of Optics and Photonics, National Central Univeristy, Chungli 32001, Taiwan
国际会议
上海
英文
2018-2021
2011-10-15(万方平台首次上网日期,不代表论文的发表时间)