Design of a Hand Exoskeleton Rehabilitation Device
As stroke is one of the most morbidity disease, some rehabilitation devices has been developed to solve the problem, but there isnt maturation in hand. This paper presents a new hand exoskeleton rehabilitation device to improve the result of rehabilitation and help doctors to perfect theory about hand rehabilitation. We set 8 DOF in this three fingers device, with 2 active DOF and 1 coupled DOF in index or middle finger, 1 active DOF and 1 passive DOF in thumb. The 5 active DOF is driven by motors. The transmission consists of screw nut and multi-link mechanism. We control the device through a typical secondary control system. As this device is designed to help patients not only regain the range of flexion/extension but also complete some typical simple grasping movements. To substantiate the feasibility, an experiment has been carried out. Finally, Suggestions for improvement are given.
Exoskeleton Hand Rehabilitation
QianBI Canjun YANG
State Key Laboratory of Fluid Power and Electromechanical System, Zhejiang University,Hangzhou, Zhejiang province, China
国际会议
2011 International Conference on Mechatronics and Materials Processing(2011年机电一体化与材料加工国际会议 ICMMP)
广州
英文
1778-1783
2011-11-18(万方平台首次上网日期,不代表论文的发表时间)