会议专题

The Parameters Calibration and Error Compensation of Circular Gratings for Parallel Double-joint Coordinate Measuring Machine(CMM)

In this paper, the mathematical model with errors for parallel double-joint coordinate measuring machine (CMM) was proposed. The main factor of the impact of circular grating measurement error—radial install eccentric error—was analyzed. The error was measured and the data obtained from measurement was used to curve fitting and form the error compensation formula. Experiments show that the method of error compensation has good usability and accuracy.

Parallel Double-joint Error Model Circular Gratings Error Compensation Calibration.

Xiao Wei Zhang Xing Hua Li Bo Chen

State Key Laboratory of Precision Measuring Technology and Instruments,Tianjin University, Tianjin 300072, China

国际会议

2011 International Conference on Vibration,Structural Engineering and Measurement(2011年振动、结构工程与测量国际会议 ICVSEM 2011)

上海

英文

1899-1902

2011-10-21(万方平台首次上网日期,不代表论文的发表时间)