Study on the Near Infrared Model Development of Mixed Liquid Samples by the Algorithm of OSC-PLS
The orthogonal signal correction of Wold algorithm was introduced in this thesis, which was combined with the partial least square algorithm. Cali-bration model of five components of benzene, methylbenzene, chlorobenzene, benzaldehyde and acetophenone in the mixed liquid samples were developed and validated by the method of independent. The result indicated that the calibration models of the components of benzene, methylbenzene, chlorobenzene and ben-zaldehyde will be developed well after the 1-dimension OSC filter; while a good calibration model of the component of acetophenone will be acquired after the 2-dimension OSC filter.
Orthogonal Signal Correction Partial Least Square Calibration Model Near Infrared Spectroscopy Liquid Samples
Dong Wang Zhihong Ma Shengfeng Ye Shungeng Min
Department of Applied Chemistry, College of Science, China Agricultural University, No. 2,Yuanmingyu Beijing Research Center for Agri-food Testing and Farmland Monitoring, Beijing Research Center for I Department of Applied Chemistry, College of Science, China Agricultural University, No. 2,Yuanmingyu