Research on Acquisition System of Multichannel Weak Photocurrent
This paper make experiments and analyze with the acquisition system of multichannel weak photocurrent for Laser particle size analyzer. It is found using analog multiplexers directly switching multi-channel photocurrent will have a pulse spike. Through the analysis of photocurrent source output port, it was found that distributed capacitance will cause a pulse spike after switch closed. Then proposed the existence model of distributed capacitance in photocurrent source output port and the method of distributed capacitance discharge before switch in each channel. Experiments show that this design can significantly eliminate pulse spike, increase the speed of acquisition system.
laser particle size analyzer data acquisition photo-current source distributed capacitance pulse spike
JIANG Hai-ying MOU Xiang-yang GAO Cheng-nan MA Yu-Liang
School of Information Science and Engineering University of Jinan Jinan, China 250022
国际会议
贵阳
英文
1-4
2011-01-26(万方平台首次上网日期,不代表论文的发表时间)