Design and characterization of a tissue-mimicking phantom for ultrasonic elastography

Tissue-mimicking phantoms are essential for theinvestigation of ultrasonic elastography. A modifiedpolyacrylamide gel based tissue-mimicking phantom is developedin this paper. It shows greater stability than phantom based onphysical gel, such as gelatin and agar. The elastic andechographic properties of this phantom are determined by theconcentration of acrylamide and graphite powder respectively.Both indentation testing and transient elastography are used tocharacterize the elastic properties of this phantom. The results arealmost consistent with each other. Our work indicate thatYoungs modulus of the polyacrylamide gel based tissue-mimickingphantom is similar to that of human soft tissue, whichranged from several kilopascal to a few tens of kilopascal.

Tao Ling Qiaofeng Jin Hui Yao Hairong Zheng

Paul C ·Lauterbur Research Center for Biomedical Imaging, Institute of Biomedical and Health Engineering, Shenzhen Institute of Advanced Technology,Chinese Academy of Sciences Shenzhen, China

国际会议

The 4th International Conference on Bioinformatics and Biomedical Engineering(第四届IEEE生物信息与生物医学工程国际会议 iCBBE 2010)

成都

英文

1-4

2010-06-18(万方平台首次上网日期,不代表论文的发表时间)