Cross-talk Calibrating Based on Correlation Analysis in High-throughput Genome Sequencing
The images, in high-throughput genome sequencing of sequencing by synthesize, consist of A, C, G, T base fluorescence images, and the fluorophore emission spectra partially overlap, which causes cross-talk problem. This paper presents a method of correlation analysis, construct the calibration matrix to eliminate interference of the fluorophore. The experiment result shows the method provided in this article is effective to correct cross-talk influence, and get the deconvolution of the clusters signal.
Bing-gang Ye Xiao-ming Wu Li Jing-xiang Zhou Yan Wang De-peng
School of Bioscience and Bioengineering, South China University of Technology Guangzhou, China Shenzhen BGI Shenzhen, China
国际会议
成都
英文
1-4
2010-06-18(万方平台首次上网日期,不代表论文的发表时间)