会议专题

Cross-talk Calibrating Based on Correlation Analysis in High-throughput Genome Sequencing

The images, in high-throughput genome sequencing of sequencing by synthesize, consist of A, C, G, T base fluorescence images, and the fluorophore emission spectra partially overlap, which causes cross-talk problem. This paper presents a method of correlation analysis, construct the calibration matrix to eliminate interference of the fluorophore. The experiment result shows the method provided in this article is effective to correct cross-talk influence, and get the deconvolution of the clusters signal.

Bing-gang Ye Xiao-ming Wu Li Jing-xiang Zhou Yan Wang De-peng

School of Bioscience and Bioengineering, South China University of Technology Guangzhou, China Shenzhen BGI Shenzhen, China

国际会议

The 4th International Conference on Bioinformatics and Biomedical Engineering(第四届IEEE生物信息与生物医学工程国际会议 iCBBE 2010)

成都

英文

1-4

2010-06-18(万方平台首次上网日期,不代表论文的发表时间)