Double-Column Capillary GC Determination of the Organophoshate Pesticide Residues in Tea
In this research, a determination of 6 organophosphorus pesticide residues in the tea has been established by capillary gas chromatography. After rotary steam and condensation, the sample is filtered through 0.2μm membrane and determined by gas chromatography FPD detector. The peak time of six pesticide ranges from 9.077min to 32.420min. The average recovery ranges between 90.5% and 110.6%, RSD, between 13% and 8.1%, detection limits, between 0.0066μg·kg-1 and 0.0085μg·kg-1, and the correlation coefficient of the standard curve, between 0.99848 and 0.99988 in the concentration range of O.O1-1.Omg·L-1.
Capillary gas chromatography(GC) organophOsphorus pesticide residues iea
Xiong Jia Lian Min Yang Yi-jian Chen Tian-cheng Liu Hong-Bin Wang
School of Chemistry and Biotechnology ,Yunnan University of Nationalities KunMing, China
国际会议
昆明
英文
323-326
2010-10-17(万方平台首次上网日期,不代表论文的发表时间)