会议专题

Research of Correlation on Calibrating and Comparing Equipments

For high precision measurement, this paper introduces the correlation principle on calibrating and comparing equipments. Frequency comparing equipments’ sensitivity, phase comparing system and double mixing time difference (DMTD) system’s self-calibration can be realized by using auto-correlation principle. Calibration of measuring equipments can be done by using cross-correlation principle. Study and application in time and frequency domain measurement is expatiated by instances using correlation principle. From the measured data, it proves that the method improves the stability test of reliability.

metrology comparing equipment calibration correlation self-calibration

Shi yongbin Dai Jun Wang Yuzhen

Unit 92493 of PLA,Huludao 125000,China Unit 91550 of PLA,Dalian 116023,China

国际会议

2011 10th International Conference on Electronic Measurement & Instruments(第十届电子测量与仪器国际会议 ICEMI2011)

成都

英文

1360-1363

2011-08-16(万方平台首次上网日期,不代表论文的发表时间)