会议专题

Reliability Evaluation and Improvement of Deep Brain Stimulator Based on Accelerated Flexing Test

Reliability of a deep brain stimulator is very important. In this paper, the reliability performance of the extension and lead of a deep brain stimulator was evaluated based on an accelerated flexing test, and through a weakness detection, segment level improvement and assembly level improvement procedure, failure frequency was reduced and bending count till failure values got significant increased for the weak points of the extension and lead. The deep brain stimulator is now under clinical trial, and the results showed that the method and improvement strategy is helpful to reliability design and process improvement.

deep brain stimulation long-term reliability accelerated flexing test

Changqing Jiang Xiongwei Wen Hongwei Hao Luming Li

School of Aerospace Tsinghua University Beijing, China

国际会议

2011 9th International Conference on Reliability,Maintainability and Safety(第九届国际可靠性、维修性、安全性会议 ICRMS2011)

贵阳

英文

50-53