会议专题

The error of DL-6A Resistance Vacuum Gauge

This paper presents the error of resistance (Pirani ) vacuum gauge. The measurement range of mis kind of vacuum gauge is 10-10~5Pa. When the vacuum degree is in the range of (10-5)?03Pa,the error is about 10%. However,provided that the vacuum degree is less than l0Pa or more than 5003Pa,the error can reach to 35%.

WANG Xun

Department of Electronics,Peking University,Beijing,100871

国际会议

10th International Conference on Vacuum Metallurgy and Surface Engineering,Vacuum Engineering Conference 2011 and Vacuum Consultancy Workshop 2011(第十届国际真空冶金与表面工程学术会议、2011年真空工程学术会议、2011年真空咨询工作会议)

沈阳

英文

385-388