会议专题

Construction of Balance Ability Evaluating Equipment Based on Passive Motion

This thesis evaluates the level of body balance ability, by adding a disturbance of horizontal displacement to the tasters feet, to collect the testee adjusting balance feedback data. It makes a contrast between the balance adjusting process curve, produced on condition that the feet get a disturbance, and Second-order system step response curve. It proposes the settling time as evaluation criteria of human balance. The design includes a body center of gravity projection force plate, consisting of three pullpressure sensors, and adopts sixdegree-of-freedom motion platform to supply a horizontal direction movement for the testee. The device includes an embed system, which owns the function of collecting analog signal, AM) conversion, coordinate calculation and sending data to PC. The host computer software, which is exploited by LabVIEW, adopts to display gravity position and save data. The off-line calculation software could handle data saved in the files with the form of MATLAB data, and calculate settling time Ts automatically, which is designed by the mixed program techniques of LabVIEW and MATLAB. Finally, hundreds of people were measured by the system, and after the off-line analysis, the thesis acquired a standard balance settling time range of normal citizens.

balance ability force plate settling time LabVIEW MATLAB motion platform

Jinzhuang Xiao Kun Liu Hongrai Wang Peng Xiong

Hebci University, Baoding, hebei, 071000, China

国际会议

2011 Fourth International Conference on Intelligent Computation Technology and Automation(2011年第四届智能计算技术与自动化国际会议 ICICTA 2011)

深圳

英文

612-615

2011-03-28(万方平台首次上网日期,不代表论文的发表时间)