The Error Analysis of Automated Biochemical Analyzer
In order to find out the relationships between the four error sources, namely, locating error, optics system fluctuations, sampling error and temperature fluctuations, and the accuracy of the instrument, we analyzed each relationship of them, and found that the effect of locating error is so tiny that it can be ignored, finally established the total error model and the model of the allowed noise of the optics system of the instrument. Analysis and the models may supply a reliable theory evidence for the improvement of the instrument and the designing of similar instruments.
biochemical analyzer optical absorbance error analysis optics system fluctuations