Design of sector multi-channel detecting system for magnetic induction tomography

Magnetic induction tomography (MIT) is a new method to get the information of bio-impedance distribution. In this paper, we propose a sector multi-channel detection system for MIT to get the impedance information in the detecting area. In the system, the exciting coil and detecting coils were designed separately to enhance the excitation magnetic field intensity, and improve the detecting resolution. Moreover, sector design provides a new way to reduce the illconditioned system of equations for image reconstruction. The system has been tested in air field and physiological saline, and the experimental results of △B/B indicates that the proposed system is efficient for impedance detecting, and this technique could potentially improve the performance of MIT system.
magnetic induction tomography (MIT) biompedance sector multi-channel detection system
Qiang Du Baodong Bai Li Ke Xiao Lin
Institute of Biomedical and Electromagnetic Engineering Shenyang University of Technology Shenyang, Liaoning 110870, China
国际会议
太原
英文
129-132
2010-10-22(万方平台首次上网日期,不代表论文的发表时间)