会议专题

New Design of Double-core QCM System

QCM(Quartz Crystal Microbalance) is a piezo electricity sensor with nanogram precision. However twice measurements of no-load and a load are necessary in the tradition, which will result in frequency errors when environments(such as temperature and humidity) changes. This paper affords a new design of double-core QCM for solving the timedelay problem in the QCM traditional measuring method. Reference QCM is introduced, QCM signal drive and acquisition circuit based on correlation theory is designed and frequency characteristic curve of double-core QCM is analyzed.

QCM double-core correlation theory frequency characteristic curve PACC:8280T

Haifeng Jiang Yuming Bo

Institute of Automation, Nanjing University of Science and Technology 200 Xiao Ling Wei Street Nanjing 210094,China

国际会议

2010 International Conference on Microwave and Millimeter Wave Technology(2010国际微波与毫米波技术会议 ICMMT2010)

成都

英文

921-924

2010-05-08(万方平台首次上网日期,不代表论文的发表时间)