New Calibration Method for the High-precision and Full-immersion Marine Thermometer
According to the changes in temperature and other parameters in the marine environment, we research the measurement accuracy impact factors of the fullimmersion marine thermometer and its calibration method. Through laboratory calibration data of the conductivity, temperature and depth (CTD) instrument to amend the zero and temperature drift of the thermometer. At the same time we compare the thermometer sea-trial data to the MTL (Import) instrument measurement. After the trial and data processing, we get that the temperature error of the full-immersion thermometer is less than 2mK.
Ocean temperature measurement calibration Sea trial comparing measurement
Yu Keguang Xu Xing Luo Xianhu Zhou Hangxia Li Qing
Zhejiang Water Conservancy and Hydropower College,Hangzhou 310018 Guangzhou Marine Geological Survey,Guangzhou 510760 China Jiliang University,Hangzhou 310018
国际会议
厦门
英文
1368-1373
2010-05-22(万方平台首次上网日期,不代表论文的发表时间)