ROBUSTESS OF THE CALUX BIOASSAY: STATISTICAL AALYSIS OF THE BETWEE-WELL VARIABILITY FOR THE H1L6.1C3 MOUSE HEPATOMA CELL LINE
The variation of the CALUX response between different wells on a 96-well plate was investigated for the H1L6.1c3 mouse hepatoma cell line. Analysis of 6 plates, spiked with 1.5 pg/μL of a 2, 3, 7, 8 -TCDD standard solution, on 3 different days indicated significant differences in RLU values between the wells. The mean RLU values were significantly lower (p = 1.43E-12) in the outer circle of the plate (i.e. rows A and H, and columns 1 and 12) and there were significant differences found between both columns and rows in a two-way ANOVA (p < 0.001). Even after rejection of the outer wells of the plate, significant differences were still found between the rows, columns (both p-values < 0.001) and circles (p = 0.002). Therefore, it is important to pay attention to the spatial distribution of the samples, standards and blanks on the plate and to take into account the between-well variability on a 96-well plate.