Development of a Calculable Current Transformer with Wide Ratio Error of - 9 % to + 11 % and its Application
Current transformers (CT) are used for precise high current measurement in power industries. A calculable CT with nominal ratio error of –9 % to +11 % has been developed on basis of theoretical calculation of ratio error by the number of windings. The developed CT with wide ratio error (WRE CT) can be used to evaluate the linearity and accuracy of the CT comparator by comparing both the theoretical and experimental values of the WRE CT. The WRE CT has been applied for the on-site calibration and correction of the CT comparator with wide ratio error range of – 9 % to + 11 %.
calculable current transformer (CT) turn ratio error self ratio error CT with wide ratio error CT comparator on-site calibration
Sung-Won Kwon Jae Kap Jung Sang-Hwa Lee Mun-Seog Kim
Korea Research Institute of Standards and Science, Daejon 305-340, Korea