A New Method to Improve the Precision of SO2 Measurement in Atmosphere
The thesis discusses the status quo and existent problems of SO2 measurement, explores sensor mechanism of measurement precision of SO2, and deduces the mathematical measurement model of SO2 according to Lambert-Beer Law. Aiming at the nonlinearity problem existing in the mathematical model, it employs Taylors series expansion to resolve the problem. In addition, it designs the gas room of double optical paths to measure SO2 by infrared, which utilizes four detectors to eliminate the impact of CO2 and disturbed noise by receiving signals. Finally, through the analysis of the experimental data, it proves that this method has such features as easy implementation, high precision, strong anti-jamming capability, and real-time and online measurement of SO2 required in the industry field.